Catalog Number 4C8723
Medical Device Problem Codes Disconnection (1171); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 01/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the tubing of a clearlink system y-type blood/solution set separated from the spike.This was identified during priming of a blood bag which resulted in a leak.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
Additional information was added to h3, h4 and h6.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photograph was performed using the naked eye which observed the tubing was separated from the spike.Due to the nature of the sample, no additional testing could be performed.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Common Device Name
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11269469
Report Number1416980-2021-00352
Device Sequence Number3302735
Product Code BRZ
UDI-Device Identifier00085412049076
UDI-Public(01)00085412049076
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K993120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4C8723
Device Lot NumberR20F22155
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/11/2021
Supplement Date Received by Manufacturer03/04/2021
Initial Report FDA Received Date02/02/2021
Supplement Report FDA Received Date03/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-