| Catalog Number |
4C8723 |
| Medical Device Problem Codes |
Disconnection (1171); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing of a clearlink system y-type blood/solution set separated from the spike.This was identified during priming of a blood bag which resulted in a leak.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h3, h4 and h6.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection of the photograph was performed using the naked eye which observed the tubing was separated from the spike.Due to the nature of the sample, no additional testing could be performed.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|