|
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from merge healthcare and other vendors systems including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.Merge cardio is intended to allow users to review diagnostic and non-diagnostic quality images, annotate studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physician generated clinical reports (via structure reporting and template based tools), and to store this information in a database.On 1/15/2021, a customer contacted merge healthcare to report that the left atrial internal diameter measurement was displaying incorrectly on the cardio clinical report.This mapping string was updated and the issue was confirmed to be resolved.No further action is necessary at this time.This has the potential to delay patient treatment and/or diagnosis.There have been no reports of patient injury or harm as a result of this issue.Reference complaint (b)(4).
|