LIVANOVA CANADA CORP. CROWN PRT PERICARDIAL HEART VALVE; TISSUE HEART VALVE Back to Search Results
Model Number CNA21
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code Aortic Regurgitation (1716)
Date of Event 10/17/2020
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer received information on a patient, implanted with a crown prt valve, model cna21, on (b)(6) 2020 that was re-admitted to the hospital on (b)(6) with heart failure/aortic regurgitation.The transthoracic echocardiogram (toe) showed a leaflet dysfunction.Specifically, it was reported that there appears to be restricted motion of one of the valve cusps.There was no thrombus/pannus identified as root cause of the stuck leaflet.The patient was treated with valve in valve (tavi) on (b)(6) 2020 and the patient received 23mm medtronic evolut pro.As reported, in theatre on the toe that one leaflet was a bit strange but was mild aortic regurgitation at the time.
 
Additional Manufacturer Narrative
The manufacturing and material records for the crown prt aortic pericardial heart valve, model#: cna21, s/n#: (b)(6) , as they pertain to the reported event, were retrieved and reviewed by quality engineering at livanova canada corp.A review of the function test (open/close valve) at the time of manufacture and release was also perform.The results confirmed that this valve satisfied all material, visual, and performance standards required for a crown prt aortic pericardial heart valve at the time of manufacture and release.The manufacturer attempted to retrieve additional information on the event, but no further details has been received to date.Since the device was not explanted, no further investigation is possible at this time.Based on the information available, it is not possible to draw a definitive conclusion on the reported event.However, per the document review performed, no manufacturing deficits have been identified.Should further information become available in the future, the manufacturer will provide an update to this reporting activity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROWN PRT PERICARDIAL HEART VALVE
Common Device Name
TISSUE HEART VALVE
Manufacturer (Section D)
LIVANOVA CANADA CORP.
5005 north fraser way
burnaby, bc
MDR Report Key11270531
Report Number1718850-2021-01011
Device Sequence Number17018762
Product Code LWR
UDI-Device Identifier00896208000511
UDI-Public(01)00896208000511(240)CNA21(17)210930
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2021
Report Date (Section F) 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model NumberCNA21
Device Catalogue NumberCNA21
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/01/2021
Device Age4 YR
Event Location Hospital
Date Report to Manufacturer01/04/2021
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/30/2021
Initial Report FDA Received Date02/02/2021
Supplement Report FDA Received Date04/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-