Model Number 71360758
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during set up or inspection, the nurse staff cross threaded the handle into the trial shell, breaking both devices.S+n back up devices were used in order to complete the procedure.No surgical delay or injury to the patient was reported.
 
Additional Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the state failure mode.The threaded area of the device is heavily damaged rendering it inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 ACET TRIAL SHELL 58MM
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11270585
Report Number1020279-2021-00868
Device Sequence Number11638393
Product Code JDG
UDI-Device Identifier03596010599117
UDI-Public03596010599117
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71360758
Device Catalogue Number71360758
Device Lot Number08ASA0111
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2021
Initial Date Received by Manufacturer 01/11/2021
Supplement Date Received by Manufacturer06/12/2021
Initial Report FDA Received Date02/02/2021
Supplement Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-