SMITH & NEPHEW, INC. TAN/FAN PERC DRILL GUIDE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631021
Medical Device Problem Codes Connection Problem (2900); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during surgery, prongs that hold nail of the percutaneous drill guide device are bent and won't release the nail properly.The procedure was completed without delay using the same device.No other complications were reported.
 
Additional Manufacturer Narrative
Results of investigation: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.A visual inspection of the returned device confirmed the stated failure mode.The device is bent rendering it inoperable.The device was manufactured in 2012 and shows signs of extensive use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAN/FAN PERC DRILL GUIDE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11270630
Report Number1020279-2021-00872
Device Sequence Number3502277
Product Code HSB
UDI-Device Identifier03596010604873
UDI-Public03596010604873
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71631021
Device Catalogue Number71631021
Device Lot Number12JSP0003
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 01/12/2021
Supplement Date Received by Manufacturer03/08/2021
Initial Report FDA Received Date02/02/2021
Supplement Report FDA Received Date03/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-