Model Number 5442147
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) regarding a patient spinal therapy for spinal fracture.It was reported that the crosslink was broken during procedure.There was patient involved in the event and no further complications were reported.Additional information was received on (b)(6) 2021: the conical protrusion of the crosslink screw plug (the clamping rod) was broken and fell off.
 
Additional Manufacturer Narrative
H3: product analysis #(b)(4):part # 5442147 lot # 0743396w visual optical functional visual and optical inspection did not reveal any cracks or deformation to the crosslink.No signs of inner or outer thread damage.The threaded portion of each set screw has been sheared off and one is missing.Functional inspection of the center nut confirmed the nut was able to be tightened and loosened without any issue.This type of damage is consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key11270850
Report Number1030489-2021-00125
Device Sequence Number16289824
Product Code KWP
UDI-Device Identifier00613994588487
UDI-Public00613994588487
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K091974
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 07/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5442147
Device Catalogue Number5442147
Device Lot Number0743396W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2021
Supplement Date Received by Manufacturer06/10/2021
Initial Report FDA Received Date02/03/2021
Supplement Report FDA Received Date07/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-