|
SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 3-4 12MM; PROSTHESIS, KNEE, PATELLOFEMTIB, SEMI-CONSTRAIND, CEMENTD, POLY/METAL/POLY
|
Back to Search Results |
|
| Model Number |
71453174 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an unspecified surgery the legion ps size 3-4mm, 12mm xlpe insert was not locking properly upon usual impaction methods.The procedure was completed with less than or 30 mins delay using a smith and nephew back-up device.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection confirms the locking mechanism has damage due to the attempts to insert the implant.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|