Model Number 5540230
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare professional via field representative regarding patient with lumbar disc herniation involved in posterior thoracolumbar surgery/ internal fixation with screw rod procedure.It was reported that at intra-op, when to broke the set screw tail, the tool was broken in the set screw.After replacing the set screw with a new one, it was used normally.There was no fragments left in the patient's body.Product was never implanted, replaced with medtronic product.Event was associated with the patient.There were no patient symptoms or complications reported as a result of the event.There was no additional surgery/treatment performed as a result of the event.Patient is alive and no injury.2021-jan-19, received additional information that event occurred during implantation of screw in the patient.Tool/driver was replaced with another one during the procedure.2021-jan-28, received additional information that driver was discarded after surgery.Set screw did not break.Setscrew was unusable.
 
Additional Manufacturer Narrative
Device evaluated summary- visual and macroscopic inspection confirmed the threads of the set screw have been damaged.The thread crest and flank damage appears to have initiated at the start of the thread, and is consistent around the damaged portion of the thread.The female torx of the screw has what appears to be the tip of a driver broken off and jammed inside.This type of damage is consistent with overload of the set screw threads during construct assembly.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key11272322
Report Number1030489-2021-00130
Device Sequence Number584531
Product Code KWP
UDI-Device Identifier00643169058750
UDI-Public00643169058750
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113174
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5540230
Device Catalogue Number5540230
Device Lot Number0753834W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2021
Supplement Date Received by Manufacturer06/02/2021
Initial Report FDA Received Date02/03/2021
Supplement Report FDA Received Date06/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
-
-