| Model Number |
TMICL12.6 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm tmicl12.6, -10.50/1.0/075 (sphere/cylinder/axis), implantable collamer lens into the patient's left eye (os) on (b)(6) 2020.The lens was exchanged with a for a shorter length lens due to excessive vault on (b)(6) 2021.This resolved the problem.Reportedly, "lens too big for patient's eye.Needed replacement.".
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h3 - device evaluation: lens was returned in micro-centrifuge vial, in liquid.Visual inspection found haptic torn.Claim# (b)(4).
|
| |
|