Model Number N/A |
Device Problem
Temperature Problem (3022)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 01/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit and replaced the helium assembly, mufflers, and the drive regulator.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) was overheating.It is unknown the circumstances under which the event occurred.It is also unknown if there was patient involvement.However, there was no adverse event reported.
|
|
Event Description
|
It was reported that during a routine check the cardiosave intra-aortic balloon pump (iabp) was overheating.There was no patient involvement, thus was no adverse event reported.
|
|
Manufacturer Narrative
|
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was reviewed and no non-conformances related to the reported event were noted.
|
|