• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number 10371
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Perforation (2513); Pericardial Effusion (3271)
Date of Event 01/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a perforation and a pericardial effusion occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman access system (was) was positioned and a 21mm watchman laa closure device & delivery system (wds) were used.The device was deployed and one partial recapture was required.The device met release criteria and was released.Post-case transesophageal echo (tee) sweeps did not show any effusion.Two hours post-implant, a pericardial effusion was noted.The patient went to the operating room and had a stitch placed at the descending end of the laa.There was a small tear from the feet of the watchman device.The patient is reported to be doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11274559
Report Number2134265-2021-00999
Device Sequence Number14522972
Product Code NGV
UDI-Device Identifier08714729838227
UDI-Public08714729838227
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/10/2022
Device Model Number10371
Device Catalogue Number10371
Device Lot Number0024751609
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/14/2021
Initial Report FDA Received Date02/03/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-