WRIGHTS LANE SYNTHES USA PRODUCTS LLC 10MM/125 DEG TI CANN TFNA 380MM/LEFT - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 04.037.029S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Malunion of Bone (4529)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Manufacturing location: (b)(4), manufacturing date: 09-dec-2019, expiration date: 31-oct-2029, part number: 04.037.029s, 10mm/125 deg ti cann tfna 380mm/left¿ sterile, lot number: 27p9674 (sterile), lot quantity: (b)(4).Production order traveler met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, in-process / inspect dimensional / final met all inspection acceptance criteria apart from the one piece noted.Inspection sheet, tfna assembly inspection met all inspection acceptance criteria.Packaging label log (pll) was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn (b)(4) supplied by ethicon ((b)(4)) was reviewed and determined to be conforming.This lot met all dimensional, visual, sterilization and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component parts reviewed: component parts were not reviewed as the reported complaint condition of ¿removal due to malreduction¿ does not indicate breakage of the nail or any of its components.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021 the patient underwent a removal due to the malreduction of the original surgery.The original surgery was done at the end of (b)(6) 2020.The trochanteric fixation nail advanced (tfna) femoral nail, femoral neck screws, perforated and 5.0mm ti locking screw were removed successfully.The surgery was completed successfully with no patient consequences.This report is for one (1) 10mm/125 deg ti cann tfna 380mm/left - sterile.This is report 1 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM/125 DEG TI CANN TFNA 380MM/LEFT - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11274961
Report Number2939274-2021-00666
Device Sequence Number3477147
Product Code HSB
UDI-Device Identifier10886982095879
UDI-Public10886982095879
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K131548
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.037.029S
Device Catalogue Number04.037.029S
Device Lot Number27P9674
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/15/2021
Initial Report FDA Received Date02/03/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-