• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA FENESTRATED SCREW 95MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA FENESTRATED SCREW 95MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 04.038.195S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Malunion of Bone (4529)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021 the patient underwent a removal due to the malreduction of the original surgery.The original surgery was done at the end of (b)(6) 2020.The trochanteric fixation nail advanced (tfna) femoral nail, femoral neck screws, perforated and 5.0mm ti locking screw were removed successfully.The surgery was completed successfully with no patient consequences.This report is for one (1) tfna fenestrated screw 95mm - sterile.This is report 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A1: patient identifier: (b)(6).Part number: 04.038m195sp, lot number: 50p1647, part manufacture date: march 30, 2020, manufacturing location: elmira, dhr record review: a review of the device history record revealed, no complaint related anomalies.The device history record shows this lot of tfna fenestrated screw 95mm, sterile product was processed through the normal manufacturing.And inspection operations with no rework nor nonconformities noted.This lot was shipped to monument and assigned as part number 04.038.195s for final packaging.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented, during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed, this lot met all specifications with no nonconformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented, during the manufacture that would contribute to this complaint condition.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented, during the manufacture that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined, that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained, that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA FENESTRATED SCREW 95MM - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key11275109
Report Number2939274-2021-00667
Device Sequence Number3477148
Product Code HSB
UDI-Device Identifier10886982098757
UDI-Public10886982098757
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.038.195S
Device Catalogue Number04.038.195S
Device Lot Number50P1647
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/15/2021
Supplement Date Received by Manufacturer03/04/2021
Initial Report FDA Received Date02/03/2021
Supplement Report FDA Received Date03/09/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-