• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 52 DISCOVERY RAD-5V PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 52 DISCOVERY RAD-5V PULSE OXIMETER Back to Search Results
Model Number 26819
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.
 
Event or Problem Description
The customer reported the device doesn't display measurements correctly.No patient impact or consequences were reported.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: other, other text: the returned device was evaluated.External visual inspection found the outer shell is cracked.When powered on not all led segments illuminate when obtaining measurements.Internal inspection found contamination damage on the system circuit board that prevents some of the led segments from lighting up.A service history record review reveals that this unit was in the field for over seven (7) months with no previous reported issues related to this reported event.
 
Event or Problem Description
The customer reported the device doesn't display measurements correctly.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-5V PULSE OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 52 DISCOVERY
52 discovery
irvine CA 92618
MDR Report Key11276443
Report Number3011353843-2021-00017
Device Sequence Number4228758
Product Code DQA
UDI-Device Identifier00843997005524
UDI-Public00843997005524
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K033998
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26819
Device Catalogue Number9197
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/13/2021
Supplement Date Received by Manufacturer02/22/2021
Initial Report FDA Received Date02/03/2021
Supplement Report FDA Received Date03/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-