| Model Number |
26819 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product has been returned to the local facility but has not yet been received at the main office for evaluation.Once returned and investigated, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported the device doesn't display measurements correctly.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: other, other text: the returned device was evaluated.External visual inspection found the outer shell is cracked.When powered on not all led segments illuminate when obtaining measurements.Internal inspection found contamination damage on the system circuit board that prevents some of the led segments from lighting up.A service history record review reveals that this unit was in the field for over seven (7) months with no previous reported issues related to this reported event.
|
| |
|
Event or Problem Description
|
|
The customer reported the device doesn't display measurements correctly.No patient impact or consequences were reported.
|
| |
|