• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FEM LT SZ 6 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FEM LT SZ 6 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Model Number 1504-00-126
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Primary operative notes (b)(6) 2017 indicate the patient received a left total knee replacement due to end stage osteoarthritis.The patella was resurfaced, and depuy cement was utilized x2.The surgery was completed without indication of complication by the surgeon.Revision operative notes (b)(6) 2019 indicate the patient received a left total knee revision due to mechanical loosening.Upon entering the joint, the femoral component was noted to be loose at the cement to implant interface.The femoral, tibial and insert components were revised with competitor product.The patella was left unrevised, no indication of deficiency with the patella, insert or tibial base.The surgery was completed without indication of complication by the surgeon.Doi: (b)(6) 2017.Dor: (b)(6) 2019.Left knee.
 
Additional Manufacturer Narrative
Product complaint #
=
> pc-(b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM LT SZ 6 NAR CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11276878
Report Number1818910-2021-02387
Device Sequence Number2590018
Product Code JWH
UDI-Device Identifier10603295041153
UDI-Public10603295041153
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1504-00-126
Device Catalogue Number150400126
Device Lot NumberHD1473
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/22/2021
Supplement Date Received by Manufacturer03/08/2021
Initial Report FDA Received Date02/03/2021
Supplement Report FDA Received Date03/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FB INSRT SZ 6 6MM; ATTUNE CR FEM LT SZ 6 NAR CEM; ATTUNE FB TIB BASE SZ 6 CEM; ATTUNE MEDIAL DOME PAT 32MM; SMARTSET HV BONE CEMENT 40G; SMARTSET HV BONE CEMENT 40G; ATTUNE CR FB INSRT SZ 6 6MM; ATTUNE CR FEM LT SZ 6 NAR CEM; ATTUNE FB TIB BASE SZ 6 CEM; ATTUNE MEDIAL DOME PAT 32MM; SMARTSET HV BONE CEMENT 40G; SMARTSET HV BONE CEMENT 40G
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient Weight68
-
-