Model Number ECR45M
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unknown.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Attempts are being made to retrieve the device.To date, no additional information has been received and no device has been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that during the endoscopic non-wedge resection surgery, when severing right upper lobe artery tissue on the second firing, after fired, noted some staples malformed with shape of straight leg.Used suture to oversew.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 03/10/2021.D4: batch # u5ck9c.H6: component code = others (g07).Device analysis: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that one ecr45m reload was received with no apparent damage and fully fired.No functional test could be performed due to the condition of the reload.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: for the malformed staple, tissue thickness should be carefully evaluated before firing any stapler.Holding the jaws in place for 15 seconds after closing and prior to firing may result in better compression and staple formation.Refer to echelon reload product codes chart for tissue compression requirement (closed staple height).Event described could not be confirmed as the reload was received fully fired.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAY RELOAD FOR 45MM ECHELON
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key11276944
Report Number3005075853-2021-00657
Device Sequence Number19760889
Product Code GDW
UDI-Device Identifier10705036001751
UDI-Public10705036001751
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model NumberECR45M
Device Catalogue NumberECR45M
Device Lot NumberU40D2Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2021
Initial Date Received by Manufacturer 01/15/2021
Supplement Date Received by Manufacturer02/12/2021
Initial Report FDA Received Date02/03/2021
Supplement Report FDA Received Date03/10/2021
Patient Sequence Number1
-
-