Catalog Number STT-GS-003 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that a pair new transmitter message occurred.Data was evaluated and the allegation was confirmed as a pair new transmitter message was confirmed.The probable cause was determined to be a transmitter failed error.The reported event of a pair new transmitter message is reportable based on the finding of a transmitter failed error.No injury or medical intervention was reported.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that a pair new transmitter message occurred.The product was evaluated.An external visual inspection was performed and passed.A voltage test was performed and failed due to 0 vdc.A review of the share logs was performed, and a pair new transmitter message and transmitter failed error was found within the investigation window.The allegation was confirmed.The probable cause was determined to be a transmitter failed error.No injury or medical intervention was reported.
|
|