• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 100 - AMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 100 - AMER Back to Search Results
Model Number 27001
Device Problems Device Displays Incorrect Message (2591); Charging Problem (2892)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/21/2020
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported complaint of the internal battery degraded warning alarm and an error message (sf180) related to a battery charger fault.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.(b)(4).
 
Event Description
It was reported to resmed that an astral device displayed error messages related to a power fault and charging issue and internal battery degraded warning alarm.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AU  2153
MDR Report Key11277011
MDR Text Key230227102
Report Number3007573469-2021-00234
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270019
UDI-Public(01)00619498270019(11)190516(10)1381853
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27001
Device Catalogue Number27001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/03/2021
Distributor Facility Aware Date01/08/2021
Device Age19 MO
Date Report to Manufacturer02/03/2021
Date Manufacturer Received01/08/2021
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-