| Model Number |
VSI |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
Fall (1848)
|
| Date of Event |
01/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The after treatment patient couch movement was caused by set screws on the connecting linkage to the roll motor that were loose.When the patient got up, they grabbed the patient right side of the couch and swung their body to patient's left.This motion dislodged the linkage that was loose.
|
| |
|
Event or Problem Description
|
|
The patient was sitting up on the treatment couch after treatment while the staff member was at the patient's side.When the patient stood up, the couch tilted and the patient fell off the couch.
|
| |
|