| Model Number |
7010 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation by mfr: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 43.1cm distal of the strain relief.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported kink.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 28jan2021.It was reported that shaft kink occurred.The 90% stenosed target lesion was located in the mildly tortuous and mildly calcified left coronary artery (lca).A 3.0mm x 12mm quantum maverick balloon catheter was advanced for dilatation; however, the delivery shaft was kinked.No patient complications were reported and the patient status was stable.However, returned device analysis revealed shaft break.
|
| |
|