| Model Number |
391235 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
The orsiro drug-eluting stent system was chosen for the treatment.During unpacking and removing of the stent protector a stent deformation was noticed.Thus another device was used.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the stent is severely deformed at its distal end, i.E.Several struts are bent outwards.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the lifted struts were initially crimped on the balloon.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|
| |
|