• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ECHO POR FMRL LAT NC 13X145MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ECHO POR FMRL LAT NC 13X145MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Patient Problem/Medical Problem (2688)
Date of Event 06/05/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D.O.B: unknown month and day in (b)(6).Concomitant medical devices: catalog#: 010000663 g7 pps ltd acet shell 52e lot#: 6320647; catalog#: 650-1162 delta cer fem hd 32/0mm t1 lot#: 2018080547; catalog#: 110003626 biolox delta cer lnr 32mm e lot#: 3745537.Foreign: (b)(6).At this time, it is unknown if the device will be returned for evaluation, due to location of device is unknown; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2021-00352.
 
Event or Problem Description
It was reported that approximately one month post implantation, the patient experienced a superficial skin infection.The infection was resolved with medication.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The reported event of superficial infection <30 days occurred post implantation.Superficial infections occurring <30 days from implantation are considered a procedure related complication as no device has been reported as revised.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors, such as this patient's bmi of 37.18.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations, further eliminating the implanted devices as a potential source for the reported infection.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.As device has not been indicated as revised and there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the reported infection.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHO POR FMRL LAT NC 13X145MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11280832
Report Number0001825034-2021-00351
Device Sequence Number16791912
Product Code KWA
UDI-Device Identifier00880304462922
UDI-Public(01)00880304462922(17)280301(10)147410
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K070274
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number192113
Device Lot Number147410
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/25/2021
Supplement Date Received by Manufacturer02/09/2021
Initial Report FDA Received Date02/04/2021
Supplement Report FDA Received Date02/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient Weight110
-
-