• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number 10373
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Low Blood Pressure/ Hypotension (1914)
Date of Event 01/08/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported back bleed did not occur and the procedure was cancelled.A left atrial appendage (laa) closure procedure was being performed.The patient had a thrombus on the transseptal guidewire and sheath, so both were removed and pressure was held on the right femoral vein.A small hematoma was noted so pressure was held.The left femoral vein was used, and the transseptal was successful.A double curve watchman access system (was) was positioned and a 27mm watchman laa closure device & delivery system (wds) were used.The 27mm device was under compressed, so a 30mm device was prepped.The was sheath was getting air in the sheath and would not bleed back so a new double curve was placed.The 30mm device was inserted and it was noticed the patient had a drop in blood pressure (70 systolic) and the sheath was still was not bleeding back.A large hematoma was found in the right groin.The 30mm device was removed, and the sheath was brought to the right side and protamine was given.Pressure was held to the groin, and 2 units of blood were given.The estimated blood loss from the procedure was 500-600cc and all blood loss was femoral.The patient was stable two hours post-procedure.The physician felt that the air was caused by loss of volume creating a low pressure system in the left atrium.The patient was discharged (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN ACCESS SYSTEM
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11282393
Report Number2134265-2021-00706
Device Sequence Number3440552
Product Code NGV
UDI-Device Identifier08714729838203
UDI-Public08714729838203
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/20/2021
Device Model Number10373
Device Catalogue Number10373
Device Lot Number0023114234
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/08/2021
Initial Report FDA Received Date02/04/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age87 YR
Patient Weight66
-
-