| Model Number |
2178-63-177 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that mbt revision reamers are worn down at their attachment point and will not properly attach to the modified hudson adapter and need to be replaced.Please send replacements to north carolina specialty hospital in durham, nc.No additional information is available.No surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : examination of the returned device confirmed the reported event.The investigation attributed the root cause to normal use and servicing and did not establish that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|