| Model Number |
CD3257-40Q |
| Medical Device Problem Code |
Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented remotely.Review of the transmission revealed non-sustained right ventricular over-sensing related to under-sensing on the discrimination channel.No intervention was performed.The patient's condition was unknown.
|
| |
|