It was reported that the patient's pump was malposition.The cause of the pump malposition was not identified.Review of the log file captured persistent pi (pulsatility index) events on (b)(6) 2021.Pi events occur when there are sudden and substantial changes in the pulsatility index, these events were considered routine.All pi events will cause the hm3 vad speed to drop to its low speed limit, which was set at 4700 rpm.The periodic log also captured low flow with high pi readings on (b)(6) 2020, (b)(6) 2020, (b)(6) 2020, (b)(6) 2020, and (b)(6) 2021.The estimated flow appeared to be fluctuating below the alarm threshold of the 2.5 lpm.These occur at times when pi is elevated.These seemed to be physiological in nature.The patient was stable and listed for transplant.The cause of the low flow alarms was thought to be due to the pump position and the low flow alarms resolved with fluid.
|
Manufacturer's investigation conclusion: the reported pump malposition could not be confirmed through this evaluation.It was reported by the center that the patient was experiencing low flow alarms due to pump malposition.The submitted log files contained multiple low flow alarms when the estimated flow decreased below a threshold of 2.5 lpm for 10 seconds or more.The pump speed remained within the set speed parameters for the duration of the log file, and the system appeared to function as intended.The patient remains ongoing on heartmate 3 left ventricle assist device (lvad), serial number (b)(6), and no further related events have been reported at this time.The heartmate 3 instructions for use (ifu) explains how to insert the pump in the ventricle and instructs the user to take care to avoid orienting the inlet towards the interventricular septum.Pump function will be compromised in the presence of inlet obstruction.This document also outlines the steps to reorient the pump.In addition, this ifu explains that the low flow hazard alarm is triggered when pump flow is less than 2.5 lpm, the pump has stopped, the pump is not operating properly, or the driveline is disconnected from the system controller.Changes in patient conditions can result in low flow, such as hypertension.This ifu and the heartmate 3 lvas patient handbook explain all system alarms and the recommended actions associated with them.Of note, low flow software decreases the low flow alarm threshold to 2.0 lpm.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The implant kit was shipped on (b)(6) 2019.No further information was provided.The manufacturer is closing the file on this event.
|