| Model Number |
7015 |
| Medical Device Problem Codes |
Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 51.6cm distal of the strain relief.There was contrast and blood in the inflation lumen, and blood in the guidewire lumen.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 14jan2021 it was reported that shaft kink was encountered.The 95% stenosed, target lesion was located in the mid left anterior descending artery.A 2.75mm x 15mm quantum maverick balloon catheter was advanced for dilatation.However, it was noticed that the delivery shaft was kinked.The procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separated.
|
| |
|