• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 6900P33MM
Medical Device Problem Codes Calcified (1077); Degraded (1153); Fluid/Blood Leak (1250); Gradient Increase (1270)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 12/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Bioprosthetic tissue valves can deteriorate with time and eventually fail contributing to regurgitation and/or stenosis.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants/replacements and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.In this case, calcification was indicated.Calcification plays a major role in the failure of bioprosthetic heart valves.Calcification of valves occurs as a progressive, time-dependent process.Tissue valve calcification is initiated primarily within residual cells that have been devitalized.Initial calcification deposits eventually enlarge and grow into a mass, which stiffen and weaken the tissue and thereby cause the prosthesis to malfunction.The mineralization of a biomaterial is generally enhanced at the sites of intense mechanical deformations generated by motion, such as the points of flexion in heart valves.Ultimately, the result of calcification is valve failure due to tearing or stenosis.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), mechanical stress related to the valve's hemodynamic performance, and glutaraldehyde fixation of tissue.Of these, the fixation process is a relatively minor contributor to calcification for edwards' tissue valves due to anti-calcification treatments during manufacturing.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprosthesis from calcifying.The root cause of this event was due to patient related factors.The calcification observed in this case was due to a progression of the patient's underlying valvular disease pathology combined with the patient's other underlying risk factors.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
Through clinical safety, it was reported that a patient with a 33mm 6900p mitral valve underwent a valve-in-valve procedure after an implant duration of 14 years due to severe mitral regurgitation, mild mitral stenosis, thickening, and torn leaflet secondary to calcification.The tmvr was performed with a 29mm 9600tfx transcatheter valve.There were no complications noted.The patient was transferred to the icu in stable condition post procedure.The patient was discharged on pod #4 in stable condition.
 
Additional Manufacturer Narrative
H11: corrected data: based on the additional information obtained, this event is no longer considered reportable and this correction is being submitted.
 
Event or Problem Description
Additional information received: there was no indication from the physician that surgical valve prematurely failed or malfunctioned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key11283004
Report Number2015691-2021-01120
Device Sequence Number4279684
Product Code DYE
UDI-Device Identifier00690103152678
UDI-Public00690103152678
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P860057/S01
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number6900P33MM
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/12/2021
Supplement Date Received by Manufacturer02/11/2021
Initial Report FDA Received Date02/04/2021
Supplement Report FDA Received Date02/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age79 YR
-
-