| Model Number |
IAP001 |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that when attempted to place the urinary catheter with an intra abdominal pressure monitoring device, the connection where the 2 devices connect, kept coming apart.The user puts a micro clave on the end of the connection, since the patient needed the monitoring.
|
| |
|
Event or Problem Description
|
|
It was reported that when attempted to place the urinary catheter with an intra abdominal pressure monitoring device, the connection where the 2 devices connect, kept coming apart.The user puts a micro clave on the end of the connection, since the patient needed the monitoring.Per follow up via email on 23feb2021, the only problem reported was the connection between the urine meter tubing and the intra-abdominal tubing.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be due to "tube od smaller than specified".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "indications for use: the bard® intra-abdominal pressure (iap) monitoring device is intended for the monitoring of intra-abdominal pressure via a foley urinary catheter.The measured pressures can be used as an aid in the diagnosis of intra-abdominal hypertension (iah) and the associated clinical syndrome of abdominal compartment syndrome (acs).Caution as an interpretive tool, the bard® intra-abdominal pressure monitoring device should be used along with other clinical indicators to aid the physician in the diagnosis of intra-abdominal hypertension (iah) and the associated clinical syndrome of abdominal compartment syndrome (acs).Device description: the bard® intra-abdominal pressure monitoring device is composed of a tubing set used for infusing fluid into the urinary bladder through the foley catheter sampling port.It utilizes a clamping device to occlude the urinary drainage tubing to form a fluid column through which pressure is measured.Important: the bard® intra-abdominal pressure monitoring device does not include a saline bag, pressure transducer or monitoring system.It adapts to the saline bag, pressure transducer and monitoring system used in your icu.The bard® intra-abdominal pressure monitoring device must be replaced at the time the urinary catheter and/or urinary drainage tubing is replaced.The bard® intra-abdominal pressure monitoring device is for use with bard® foley trays with ez-lok® sampling port.Contraindications ¿ not for use on pediatric patients ¿ not for use on patients with compromised bladder function warnings ¿ use only saline for pressure measurement.¿ ensure tubing fluid path is primed so there are no air bubbles.¿ ensure the iap sampling port is seated tightly to the catheter drainage tubing sampling port prior to use.Precautions ¿ single use only ¿ federal (u.S.A.) law restricts this device to sale by or on the order of a physician.¿ not made with natural rubber latex ¿ sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.¿ this is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Sterile contents: ¿ 30 cc syringe ¿ infusion and pressure transducer tubing ¿ saline bag spike & cap ¿ zeroing stopcock ¿ dead-end transducer cap ¿ proprietary drain tube clamp ¿ valve port ¿ check valves for controlling and directing flow ¿ statlock® foley device kit." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Bd has determined that this mdr was reported in error as it was found to be a duplicate of an event previously reported.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that when attempted to place the urinary catheter with an intra abdominal pressure monitoring device, the connection where the 2 devices connect, kept coming apart.The user puts a micro clave on the end of the connection, since the patient needed the monitoring.Per follow up via email on 23feb2021, the only problem reported was the connection between the urine meter tubing and the intra-abdominal tubing.
|
| |
|
Search Alerts/Recalls
|