| Model Number |
2968045 |
| Medical Device Problem Codes |
Crack (1135); Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from the healthcare professional (hcp) via the manufacturer representative regarding an event which occurred during an oblique approach interbody fusion procedure in a patient diagnosed with lumbar disc herniation.It was reported that the fusion cage was cracked, which caused the protruding and displacement of the fusion cage after the surgery, and revision surgery was performed.Product will be returned.No patient other patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: (b)(4) lot# h5445499 a piece ~3.5mmx13mm has broken off the cage in the area that sits against the inserter.A microscopic review of the fracture surface gives indications that the material was overloaded during a bending moment with torsion being applied through the inserter.The failure appears to be the result of material overload during installation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|