SMITH & NEPHEW, INC. R3 0 HOLE ACET SHELL 60MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED Back to Search Results
Model Number 71331860
Medical Device Problem Code Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that, the cup seal seemed to be damaged and it was necessary to take off the field and switch implants.It is unknown whether the event happened during surgery and if there was a patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device cup seal mechanism is damaged, rendering device inoperable.The device was manufactured in 2019 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.The potential probable causes for this event could include but not limited to a user or procedural error.Based on this investigation, the need for corrective action is not indicated.Without the actual products involved and/or device information, our investigation cannot proceed.If these devices or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Event or Problem Description
It was reported that, during unknown procedure the cup seal seemed to be damaged and it was necessary to take off the field and switch implants.The procedure was completed with a delay less than or equal to 30min using a smith-nephew back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 HOLE ACET SHELL 60MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11284610
Report Number1020279-2021-00996
Device Sequence Number11692681
Product Code MRA
UDI-Device Identifier03596010597632
UDI-Public03596010597632
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K070756
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71331860
Device Catalogue Number71331860
Device Lot Number19KW02824
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2021
Initial Date Received by Manufacturer 01/14/2021
Supplement Date Received by Manufacturer01/14/2021
06/07/2021
Initial Report FDA Received Date02/05/2021
Supplement Report FDA Received Date02/05/2021
06/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-