The full identifier is case (b)(6).The customer did not provide patient demographics such as age, date of birth, sex, weight, ethnicity or race.The access sars-cov-2 igm assay was not returned for evaluation.There were no reports of system issues at the time of the event.All assay and system verifications met specifications at the time of the event.No hardware errors or flags were reported in conjunction with the event.In conclusion, the cause of this event cannot be determined with the available information.
|
|
On (b)(6) 2021 the customer reported negative covid igm results (access sars-cov-2 igm, part number c58957 and lot number 971256) were generated on the customer's dxi (unicel dxi 800 access analyzer, part number 973100 and serial number (b)(4)) for an unknown number of patients who had positive covid igm results from an alternative platform (alternative platform information not provided).The customer did not indicate whether the results were released from the laboratory.The customer did not report a change to patient care of treatment in connection with this event.No hardware errors were reported in conjunction with this event.System performance indicators such as system check, calibration and quality control were passing within specifications at the time of the event.No issues with sample integrity were reported by the customer.Sample collection, handling and processing information such as sample type, sample volume collected, sample quality, centrifugation time and speed, storage temperature and other information was not provided by the customer.
|