• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE RP TRL BLT SZ 6-8 POR; ATTUNE CEMENTLESS INSTRUMENTS : TIBIAL KEEL PUNCH/BULLET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE RP TRL BLT SZ 6-8 POR; ATTUNE CEMENTLESS INSTRUMENTS : TIBIAL KEEL PUNCH/BULLET Back to Search Results
Model Number 2545-00-185
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
After impaction of the instrument device, the punch sheared from the edges causing the piece to break off.The punch device was carefully collected and remove from the surgery table.No pieces were left inside the patient.Xrays was performed after the surgery was completed to ensure the patient was debris-free.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information was received stating that the wing tip of the punch broke into two pieces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TRL BLT SZ 6-8 POR
Common Device Name
ATTUNE CEMENTLESS INSTRUMENTS : TIBIAL KEEL PUNCH/BULLET
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11286492
Report Number1818910-2021-02489
Device Sequence Number3842894
Product Code LRY
UDI-Device Identifier10603295419730
UDI-Public10603295419730
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2545-00-185
Device Catalogue Number254500185
Device Lot NumberNW245587
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2021
Initial Date Received by Manufacturer 01/26/2021
Supplement Date Received by Manufacturer01/26/2021
03/08/2021
Initial Report FDA Received Date02/05/2021
Supplement Report FDA Received Date02/15/2021
03/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-