Model Number 5MAXJET7BKIT-A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2021-00226.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system jet 7 reperfusion catheter (jet7), penumbra system 3max reperfusion catheter (3maxc), and non-penumbra hemostasis valve (hv).During the procedure, while advancing the 3maxc and jet7 through the hv, the 3maxc became fractured near the hub.Subsequently, the physician attempted to retract the 3maxc together with the jet7 out from the hv to remove them; however, the jet7 became fractured.Therefore, both the 3maxc and jet7 were not used for the remainder of the procedure.It is unknown how the procedure was completed or if there was any adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned jet7 confirmed that the catheter was fractured.If the jet7 is forcefully retracted against resistance, damage such as a fracture may occur.Evaluation of the returned 3maxc confirmed that the catheter was fractured.If the 3maxc is forcefully advanced at an extreme angle, damage such as a kink and subsequent fracture may occur.Stress marks were present near the fractured location indicating the fracture was likely the result of a kink.Further evaluation revealed a kink in the proximal shaft.This damage was likely incidental to the reported complaint.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 1.3005168196-2021-00226.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM JET7 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11287790
Report Number3005168196-2021-00225
Device Sequence Number3475753
Product Code NRY
UDI-Device Identifier00815948023246
UDI-Public00815948023246
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K173761
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/07/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2023
Device Model Number5MAXJET7BKIT-A
Device Catalogue Number5MAXJET7BKIT
Device Lot NumberF99994
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 01/07/2021
Supplement Date Received by Manufacturer03/22/2021
Initial Report FDA Received Date02/05/2021
Supplement Report FDA Received Date04/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age49 YR
Date Report Sent to FDA01/01/2005
-
-