| Model Number |
UE00N69 |
| Medical Device Problem Codes |
Failure to Transmit Record (1521); Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation concluded the potential root cause is a configuration issue (opal-rad side).Customer sites (site a and site b) requested a destination change during a conference call with a kmha engineering employee for vpn/forwarder services, which potentially contributed to the workflow disruption as concomitant devices.Workflow started with site a (mri modality and worklist) and expected to transfer to site b with multiple opportunities for disruption.Opal has a special transfer syntax "xferobject" utilized when transferring from opal-to-opal system.Opal-rad sending to exa utilizing the transfer syntax 'xferobject' caused zero (0) data files to be generated.The files queued in the receiver on exa causing the service to remain in a continuous transfer loop.Kmha is implementing a future code improvement for exa to ensure if the misconfigure occurs in the future, the transfer is rejected immediately.
|
| |
|
Event or Problem Description
|
|
Customer reported that studies are being received but not moving from cache.Customer reported potential delay in patient care, as the date of the study was (b)(6) 2021 and the study was not received until (b)(6) 2021.
|
| |
|