• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PRODIGY MRI IPG, 16CH, RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PRODIGY MRI IPG, 16CH, RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3772
Device Problem Material Fragmentation (1261)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for evaluation.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related mfr report: 1627487-2021-00987.It was reported the patient's scs lead extension broke inside the implantable pulse generator (ipg) header port, one of the extension contacts remained inside the header.Reportedly, the patient¿s scs system exhibited high impedance and the stimulation therapy was affected due to the breakage.The breakage was confirmed during the revision when the extension was disconnected from the ipg.The extension was replaced and the issue was resolved.The generator remained implanted in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODIGY MRI IPG, 16CH, RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key11288959
MDR Text Key230682677
Report Number3006705815-2021-00457
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017093
UDI-Public05415067017093
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/20/2022
Device Model Number3772
Device Catalogue Number3772
Device Lot NumberA000099297
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-