Model Number 2011-01-270
Medical Device Problem Codes Break (1069); Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon noticed the problem on actis reamer stabilit and saw the risk of the broken reamer during surgery.Thickness of the reamer body seams are too weak even though no event happened.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.The investigation could not verify or identify any product contribution to the reported event with the information provided.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history review: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS REAMER SZ 12
Common Device Name
REAMERS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11289437
Report Number1818910-2021-02533
Device Sequence Number3498486
Product Code MEH
UDI-Device Identifier10603295495727
UDI-Public10603295495727
Combination Product (Y/N)N
Initial Reporter CountrySZ
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2011-01-270
Device Catalogue Number201101270
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer04/13/2021
05/03/2021
05/20/2021
Initial Report FDA Received Date02/05/2021
Supplement Report FDA Received Date04/14/2021
05/04/2021
05/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-