MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
THPX470 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
10/19/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture/device migration , ptosis, double bubble deformity.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old caucasian female patient underwent a primary breast augmentation with two 470cc mentor memorygel xtra breast implant prostheses and experienced postoperative right-sided grade iv capsular contracture and bilateral ptosis postoperatively.During the consultation held on (b)(6) 2020, the patient's physician stated that the right breast appeared to be higher than the left side, and the patient experienced right double bubble deformity and device migration due to the capsular contracture.The diagnosis was confirmed via physical examination.As a result, the patient underwent bilateral mastopexy and unilateral removal and replacement with a 470cc mentor memorygel xtra breast implant (catalog #: thpx470, right serial #: (b)(4)) on (b)(6) 2020.This report is for the right-sided prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2021, the investigation on the suspected medical device was completed.Investigation summary : mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer's reference number: (b)(4).
|
| |
|