|
|
| Model Number |
71070705 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
01/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, after a tsf construct had been implanted on (b)(6) 2020 for correcting a distal radius partial growth arrest in left hand, the patient experienced two half-pin track infections proximally, which resulted in loosening of the pins and an overall increasing instability of the frame.Upon removal of construct on (b)(6) 2021, the physician also noticed 'play' while manipulating two struts including the spatial frame strut x-short and the fast fx strut x-short, which raised questioning on whether they are 'structurally sound and stable in their joints and internal mechanism'.Based on the two half-pin track infections and the two alleged struts, as well as x-rays revealing that the osteotomy site was closing and the distal fragment was deemed clinically stable enough in its corrected position, the decision was made to remove the frame and finish the latter stages of healing in a pop cast.The proximal half-pins could be pulled out of the bone without any counterclockwise turns.The patient recovery is ongoing.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms although the two provided photos confirm the complaint, however, without the requested clinical information, lab report, operative reports or the device, the clinical root cause of the reported infection cannot be determined.Per the complaint, the reported infection was treated with extra cleaning and oral antibiotics (b)(4) and the overall instability (b)(4) treatment was continued with the plaster.The reported ¿play¿ during manipulation of the spatial frame strut x-short (b)(4) and the fast fx strut x-short (b)(4) occurred during the construct removal.Although the surgeon questioned the stability of the frame, the patient¿s x-ray revealed the closing of the osteotomy site.In addition, the distal fragment was deemed clinically stable in its corrected position.As a result, the surgeon removed the frame and changed the patients to a pop cast for the latter phase of healing.Since the patient¿s recovery is ongoing, no further clinical /medical assessment is warranted at this time.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This included a review of sterilization documents which indicated that the product was sterilized according to sterilization release documentation.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Some potential probable causes for this event could include but not limited fit/sizing, alignment, surgical technique, joint laxity, traumatic injury and procedural/user error.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.A functional evaluation of the returned device does not confirm the stated failure mode.The device functions as necessary.The clinical/medical evaluation concluded that although the two provided photos confirm the complaint, however, without the requested clinical information, lab report, operative reports or the device, the clinical root cause of the reported infection cannot be determined.Per the complaint, the reported infection was treated with extra cleaning and oral antibiotics and the overall instability treatment was continued with the plaster.The reported ¿play¿ during manipulation of the spatial frame strut x-short and the fast fx strut x-short occurred during the construct removal.Although the surgeon questioned the stability of the frame, the patient¿s x-ray revealed the closing of the osteotomy site.In addition, the distal fragment was deemed clinically stable in its corrected position.As a result, the surgeon removed the frame and changed the patients to a pop cast for the latter phase of healing.Since the patient¿s recovery is ongoing, no further clinical /medical assessment is warranted at this time.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.A review of sterilization documents indicated that the product was sterilized according to sterilization release documentation.Some potential probable causes for this event could include but not limited fit/sizing, alignment, surgical technique, joint laxity, traumatic injury and procedural error.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|