DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 40CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0480-07
Device Problem Defective Component (2292)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2021
Event Type  malfunction  
Manufacturer Narrative
Event site postal code: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event Description
It was reported that during insertion a defect was noticed near the tip of the intra-aortic balloon (iab).There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint #(b)(4).
 
Event Description
It was reported that during insertion a defect was noticed near the tip of the intra-aortic balloon (iab).There was no patient harm or adverse event reported.
 
Event Description
N/a.
 
Manufacturer Narrative
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id #: (b)(4).
 
Manufacturer Narrative
Additional information: changed serial number to: (b)(6).Device evaluation: the product was returned with the membrane completely unfolded with trace amounts of blood in the balloon.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and a leak was detected on the membrane seal approximately 28.96cm from the iab tip and measuring approximately 0.076cm in length.The reported problem was most likely triggered by the leak found on the membrane rear seal.The evaluation confirms the reported problem.A non-conforming material report review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period nov-19 through jan-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 07004
MDR Report Key11293209
MDR Text Key230716902
Report Number2248146-2021-00067
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567109718
UDI-Public10607567109718
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K041281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 11/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2022
Device Model Number0684-00-0480-07
Device Catalogue Number0684-00-0475
Device Lot Number3000108623
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2021
Initial Date FDA Received02/08/2021
Supplement Dates Manufacturer Received04/01/2021
06/30/2021
11/19/2021
Supplement Dates FDA Received04/01/2021
07/20/2021
11/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
Patient Weight68 KG
-
-