EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT RSRPERICARDIAL BIOPROSTHESIS - AORTIC; HEART-VALVE, REPLACEMENT
|
Back to Search Results |
|
| Model Number |
2800 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Heart Failure/Congestive Heart Failure (4446); Pulmonary Valve Insufficiency/ Regurgitation (4452)
|
| Date of Event |
01/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Regurgitation which develops progressively over time can be due to a number of issues including patient-related factors or structural valve deterioration.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly.Regurgitation may also develop progressively if host fibrotic tissue, or pannus, grows onto the bioprosthetic valve.Pannus, a cause of nonstructural dysfunction, may interfere with the functionality of the device by restricting the leaflet motion leading to abnormal coaptation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis and/or regurgitation.The root cause of this event cannot be conclusively determined with the available information.However, the regurgitation, in this case, was most likely impacted by the progression of the patient's underlying valvular disease pathology with or without structural valve deterioration and/or nonstructural dysfunction.The subject device is not available for evaluation, as it remains implanted in the patient.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.In addition, other patient risk factors such as the patient's younger age at implant may have contributed to this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, an appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 2800 aortic valve, implanted in the pulmonary position, underwent a valve-in-valve after an implant duration of five (5), six (6) months due to regurgitation.The tpvr was performed with a 26mm 9600tfx valve.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 25mm 2800 aortic valve, implanted in the pulmonary position, underwent a valve-in-valve after an implant duration of five (5), six (6) months due to moderate-severe pulmonary insufficiency.The tpvr was performed with a 26mm 9600tfx valve without complications.On pod #1, the patient was discharged home in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: updated sections: g3, g6, and h2.H11: corrective data: corrected section: h6 (investigation findings, and investigation conclusions).
|
| |
|