DEPUY CMW - 9610921 SMARTSET MV 20G - EO; BONE CEMENT : BONE CEMENT Back to Search Results
Model Number 3122-020
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 09/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
(b)(4) medical records ad 13 jan 2021 were reviewed.On (b)(6) 2016, the patient had a left knee arthroplasty.Operative notes were not provided, but a sticker sheet with product information was.Depuy components including depuy patella and depuy cement x2 were used during the procedure.On (b)(6) 2019, the patient had a revision to address tibial component loosening.Prior to the revision, the patient was noted to have a large effusion, some laxity throughout range of motion, pain and radiographs were reported to show a small area of osteolysis at the medial tibial plateau.The revision notes provided were minimal and only noted a postoperative diagnosis of tibial component loosening, with no mention of interface, effusion, or osteolysis.Postoperative radiographs from 10/14/2019 note that on the lateral view, there is some radiolucency at the anterior flange of the femur on the lateral view, however, again at the time of surgery this component was noted be very well fixed.The sticker sheet provided show only tibial components, including insert.No mention of patella or femoral component, it would be reasonable to conclude they were not revised.Doi: (b)(6) 2016 dor: (b)(6) 2019 (left knee).
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2019, the patient underwent a left knee revision to address pain, instability, edema, osteolysis, synovitis, and tibial tray loosening at the cement to implant interface.The tibial tray and tibial insert were revised.The femoral component and patella component were retained.The patient was revised with depuy products and cement.There were no indicated intra-operative complications.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET MV 20G - EO
Common Device Name
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY CMW - 9610921
cornford rd.
blackpool IN FY4 4 QQ
UK  FY4 4QQ
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key11294432
Report Number1818910-2021-02589
Device Sequence Number17014591
Product Code LOD
UDI-Device Identifier10603295168362
UDI-Public10603295168362
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K081155
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2017
Device Model Number3122-020
Device Catalogue Number3122020
Device Lot Number8069528
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/26/2021
Supplement Date Received by Manufacturer03/24/2021
12/16/2021
02/11/2022
Initial Report FDA Received Date02/08/2021
Supplement Report FDA Received Date03/25/2021
01/11/2022
02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/18/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 8 CEM; ATTUNE MEDIAL DOME PAT 41MM; ATTUNE PS FB INSRT SZ 8 5MM; ATTUNE PS FEM LT SZ 8 CEM; SMARTSET MV 20G - EO; SMARTSET MV 40G - EO; ATTUNE FB TIB BASE SZ 8 CEM; ATTUNE MEDIAL DOME PAT 41MM; ATTUNE PS FB INSRT SZ 8 5MM; ATTUNE PS FEM LT SZ 8 CEM; SMARTSET MV 20G - EO; SMARTSET MV 40G - EO
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
-
-