Model Number 9735602 |
Device Problems
Connection Problem (2900); Adverse Event Without Identified Device or Use Problem (2993); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: 9 660651r, serial/lot #: unknown, ubd: unknown, udi#: unknown.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system for functional endoscopic spinal surgery (fess).It was reported that the software was showing that the localizer was disconnected on (b)(6) 2021 and the axiem box was not communicating.The site powered off the compact and booted it with the axiem box powered on and the communication cable between the compact and the axiem box, which resolved the issue.It was noted that the healthcare provider (hcp) and manufacturer representative were setting up for a case with the patient in the room and there was about a 15 minute delay.
|
|