• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INFOBIONIC INC. MOME KARDIA; ARRHYTHMIA DETECTOR AND ALARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INFOBIONIC INC. MOME KARDIA; ARRHYTHMIA DETECTOR AND ALARM Back to Search Results
Model Number 01854
Medical Device Problem Codes Inadequate Instructions for Healthcare Professional (1319); Improper or Incorrect Procedure or Method (2017); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient affected by the event was enrolled and was monitored in a clinical research study for 29 days.To suppress billing and usage statistics, which were not desired for this study, the demo mode for software was set.In addition to suppressing these features statistics, the demo mode also allows for the upload of simulated patient data for a particular patient day, provided no other patient data has been received for that day.The patient turned off the device on last day of the monitoring, before all stored data had been transmitted, and began the process of returning the device to the practice.The device had not yet transmitted any data for the last day when it was turned off.Three days later, a monitoring technician noticed that no data had been transmitted for the last day.Since there was no data for the last day yet available, and the practice was set to demo mode, an "upload data" button was visible.The "upload data" button allows for simulated/demo data to be uploaded for any patient day containing no other data.A software feature that allows the technician to see how much data remains to be transmitted from a patient device was not used, and the technician pressed the "upload data" button under the mistaken assumption that this would cause the real patient data to be uploaded from the patient device.Instead, simulated data was uploaded for that day, and an atrial fibrillation arrythmia event was created and reported from the simulated data.Later the device was physically received back at the practice and turned on.The device then began uploading its stored data and overwrote/removed the simulated data uploaded for the last day.The monitoring technician noticed the error when preparing the final report and notified the physician on the research team.The physician then communicated the error and provided correct diagnosis to the patient and his physician.No therapy was administered to the patient as a result of initial incorrect diagnosis.
 
Event or Problem Description
A product malfunction event was reported by healthcare professional on a phone call regarding a patient who was monitored under a clinical research study.A set of simulated patient data (demo data) had been uploaded to actual patient record data inadvertently and was interpreted as actual patient data at the time of initial diagnosis.This product malfunction caused the research study patient to be incorrectly diagnosed with atrial fibrillation when initial diagnosis was done.The patient and his physician were notified of this incorrect diagnosis.One of the monitoring technicians, while finalizing the patient report later, noticed the error and notified the physician on the research team.The physician on the research team notified the patient and his physician of the error and provided correct diagnosis.No therapy was administered to the patient based on the initial incorrect diagnosis.No patient injury or harm was resulted as an outcome of the event.The simulated data was later successfully removed and original correct data was restored intact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOME KARDIA
Common Device Name
ARRHYTHMIA DETECTOR AND ALARM
Manufacturer (Section D)
INFOBIONIC INC.
200 5th avenue
suite 4030
waltham MA 02451
MDR Report Key11296106
Report Number3010112812-2021-00001
Device Sequence Number3368796
Product Code DSI
UDI-Device IdentifierB690MK10
UDI-Public+B690MK10
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K160064
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number01854
Device Lot Number092020
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2021
Initial Report FDA Received Date02/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight71
-
-