| Model Number |
N/A |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Siemens healthcare diagnostics is investigating this event.The limitations section of the instructions for use states the following: "a definitive clinical diagnosis should not be made solely on the basis of an in vitro allergen-specific ige result.Diagnosis should be made by the physician only after all clinical and laboratory findings have been evaluated.".
|
| |
|
Event or Problem Description
|
|
The customer observed a negative (class 0) result using the immulite 2000 food panel 5 assay, reagent lot 620, which was discordant relative to a positive (class 2) repeat-testing result obtained using reagent lot 621 of the same product.There are no known reports of patient intervention or adverse health consequences due to the discordant result.
|
| |
|
Additional Manufacturer Narrative
|
|
Mdr 1219913-2021-00093 was submitted on 2021-02-08 to report a negative result from the immulite 2000 food panel 5 assay which was discordant relative to repeat testing using a different reagent lot.Siemens healthcare diagnostics has concluded the incident investigation.Customer data were reviewed.Following the observation of result disagreement, the customer began to run individual-allergen tests in addition to the food panel assay, and comparing panel-result indications to individual-allergen component results.The customer noted some cases of disagreement between these test methods.Further investigation revealed that the customer was applying an inappropriate ku/l threshold for distinguishing reactive (positive) from nonreactive (negative) results.The assay instructions for use defines the range from 0.1 to 0.34 ku/l as "very low", where the customer had interpreted these results as negative.Following this clarification, there were still some observations which demonstrated inconsistency between panel and individual-allergen results.Samples were requested for testing, but were not available.Product stability data for the food panel 5 assay were reviewed, and although some decrease in measurements was observed for this assay over time, results did not change from positive to negative; no product issue was indicated.No manufacturing issues or product problems were identified.The system is working within specifications, no further support is requested by the customer, and no further action is required.Note: coding for investigation findings and investigation conclusions has been updated in section h6.
|
| |
|
Search Alerts/Recalls
|