Model Number 0998-00-3023-53 |
Device Problem
Power Problem (3010)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge authorized dealer's field service engineer (fse) evaluated the iabp unit and was able to confirm the reported issue.To fix the issue, the fse replaced the power supply and 5000 hours maintenance package.The fse then turned on the iabp and confirmed that the power supply of the iabp was working normally (battery charging and discharging and ac power).Additional information is being requested with regard to the status of the iabp.A supplemental report will be submitted when this information is provided to us.
|
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) had power supply issues.It is unknown the circumstances under which the event occurred.It is also unknown if there was patient involvement.However, there was no adverse event reported.
|
|
Manufacturer Narrative
|
A getinge service representative reported that the getinge authorized dealer's field service engineer (fse) had performed all functional and safety checks to meet factory specifications.Unit had passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during a routine check the cs300 intra-aortic balloon pump (iabp) had power supply issues.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 12 month product complaint trend data for the period feb 2020 through jan 2021 was reviewed.There were no triggers identified for the review period.
|
|