• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Model Number 1011344-40
Medical Device Problem Codes Material Separation (1562); Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Vessel Or Plaque, Device Embedded In (1204)
Date of Event 01/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual analysis was performed on the returned device.The reported tip separation was confirmed.The stent malaposition could not be confirmed as it was based on procedural conditions and the stent was not returned.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation was unable to determine a cause for the reported difficulty.It may be possible that the stent was not fully apposed due to narrowing of the vessel.Additionally, it may be possible that due to the stent being narrowed, the tip of the acculink was unable to be pulled back through the stent resulting in separation.However, this could not be confirmed.The additional treatment to attempt to retrieve the separated tip and outcome of embedding the separated tip to the vessel wall appear to be due to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the non-calcified, non-tortuous right internal carotid artery.A 7-10x40mm acculink self-expanding stent system (sess) was advanced, and the stent was deployed.The stent had not fully apposed to the vessel wall, and it was noted that the tip had separated.A snare was attempted to remove the separated tip but was unsuccessful.An unspecified stent was to embed the tip and fix the poorly apposed acculink stent to successfully complete the procedure.There were no adverse patient sequelae and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Common Device Name
CAROTID STENT
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11298193
Report Number2024168-2021-00986
Device Sequence Number17383515
Product Code NIM
UDI-Device Identifier08717648076398
UDI-Public08717648076398
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P040012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number1011344-40
Device Catalogue Number1011344-40
Device Lot Number0082661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/18/2021
Initial Report FDA Received Date02/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient Weight100
-
-