Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified broken shell, and black particles.Weight measurement identified device was underweight.A microscopic analysis was performed which identified sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identified the thickness within specification.Based on the device analysis, the final assessment is a sharp broken on radius assessed as an unidentified (tear) opening.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|