Model Number 407201
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
During the procedure, when pulling back the syringe to flush needle, air entered the syringe.The needle was replaced, and the procedure was completed with no adverse patient consequences.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Additional Manufacturer Narrative
Two brk transseptal needles were received for evaluation.Both brk needles were flushed with the sideport opened; no leak was noted.When the sideport was closed no air retracted into the syringe.No functional anomalies were noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported needle leak remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1 TRANSSEPTAL NEEDLE, 71 CM LENGTH
Common Device Name
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key11298721
Report Number3008452825-2021-00018
Device Sequence Number729424
Product Code DRC
UDI-Device Identifier05414734205160
UDI-Public05414734205160
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K072278
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model Number407201
Device Catalogue Number407201
Device Lot Number7697661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/11/2021
Supplement Date Received by Manufacturer03/01/2021
04/29/2021
Initial Report FDA Received Date02/08/2021
Supplement Report FDA Received Date03/01/2021
04/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-