| Model Number |
407201 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the procedure, when pulling back the syringe to flush needle, air entered the syringe.The needle was replaced, and the procedure was completed with no adverse patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Two brk transseptal needles were received for evaluation.Both brk needles were flushed with the sideport opened; no leak was noted.When the sideport was closed no air retracted into the syringe.No functional anomalies were noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported needle leak remains unknown.
|
| |
|