• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD HOME SHARPS CONTAINER; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD HOME SHARPS CONTAINER; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 323487
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Oem manufacture: the manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) and the (b)(4) fda registration number has been used for the manufacture report number.Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: na.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event or Problem Description
It was reported that the bd home sharps container experienced no lid and lid difficult to close/does not close.The following information was provided by the initial reporter: material no: 323487, batch no: 0024001.Purchased a bd home sharps container and it was missing the lid to the top.Consumer stated, when she opened the new box, lid that has the "holes" in it to drop pen needles ans syringes through was broken.
 
Additional Manufacturer Narrative
H.6.Investigation: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.
 
Event or Problem Description
It was reported that the bd home sharps container experienced no lid and lid difficult to close/does not close.The following information was provided by the initial reporter: material no: 323487, batch no: 0024001.Purchased a bd home sharps container and it was missing the lid to the top.Consumer stated, when she opened the new box, lid that has the "holes" in it to drop pen needles ans syringes through was broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD HOME SHARPS CONTAINER
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
MDR Report Key11298929
Report Number2243072-2021-00270
Device Sequence Number12001841
Product Code FMI
UDI-Device Identifier00382903234875
UDI-Public00382903234875
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K943139
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number323487
Device Catalogue Number323487
Device Lot Number0024001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/19/2021
Supplement Date Received by Manufacturer04/30/2021
Initial Report FDA Received Date02/08/2021
Supplement Report FDA Received Date05/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-